University of Houston University of Houston-Clear Lake ISSO Annual Report Y2003 58-65
Development of Quantum-Cascade Laser Based Biosensor Technology
Thomas L. Harman [UHCL] / Frank K. Tittel [Rice U] / John C. Graf [NASA-JSC] / Yury Bakhirkin [UHCL]
Abstract
The principle objective of this project is the development of new types of sensitive,
selective, real-time gas sensors based on continuous wave and pulsed quantum cascade
lasers for various chemical sensing applications ranging from medical diagnostics to
monitoring spacecraft air quality. Tunable laser absorption spectroscopy in the
mid-infrared spectral region is a sensitive analytical technique for trace gas
quantification. Nitric Oxide (NO) detection from exhaled breath is of particular interest for the
diagnosis of lower airways inflammation associated with a number of lung diseases and
illnesses. Investigation focuses upon the feasibility of detecting NO in expired human breath as a potential non-invasive
medical diagnostic tool. A gas analyzer based on a CW mid-infrared quantum cascade laser
operating at ~ 5.2 mm and off-axis integrated cavity output spectroscopy (ICOS) has been
developed to measure NO concentrations in human
breath. A compact sample cell, 5.3 cm in length with a volume of < 80 cm3,
suitable for on-line and off-line measurements during a single breath cycle, has been
designed and evaluated. A noise-equivalent (SNR = 1) sensitivity of 10 parts-per-billion
by volume (ppbv) of NO was achieved. The
combination of off-axis integrated cavity output spectroscopy with wavelength modulation
resulted in a 2 ppbv noise-equivalent sensitivity.
Introduction
Nitric oxide detection is important for a number of
applications, such as atmospheric pollution monitoring,1 vehicle exhaust
control,2,3 and non-invasive medical diagnostics of various lung diseases by
means of NO concentration
measurements in expired breath.4,5 In fact, exhaled breath contains a large
number of molecular species at ultra-low concentrations (ppbv), some of which are possible
biomarkers for a variety of human illnesses and diseases. Trace gas quantification in
breath is inherently non-invasive and, in some cases, can obviate the need for invasive
surgical procedures, such as biopsies. NO is involved in many biochemical processes and performs important
functions in human physiology, such as immune reactions, regulation of platelet function,
and neurotransmission.6-13 In 1998, the Nobel Prize in Physiology and Medicine
recognized discoveries concerning the role of NO in the cardiovascular system.
Concentrations of oral exhaled nitric oxide (eNO) originating from the lower airways in the absence of asthma and acute respiratory illnesses typically vary from 5 to 20 ppbv.9,10 Nasal nitric oxide (nNO) originating from the nasal cavity has higher concentrations in the range of 40 to 200 ppbv.9,10 Humming, a voice maneuver, has been reported to produce 200 to 500 ppbv of NO in expired breath.6,8
Concentrations of eNO are reduced below typical levels for patients with cystic fibrosis12 and for smokers.13 Elevated eNO concentrations in expired breath are principally associated with inflammation of the lower airways. One of the most investigated inflammatory processes is asthma. Asthma is a chronic inflammatory disorder of the lower airways, which results in increased levels of eNO in human exhaled air.9 Noninvasive diagnostics of asthma are of great importance since it afflicts millions of adults and children worldwide. Early diagnosis and intervention in asthma cases can lead to a significant reduction in hospitalizations. Typical asthmatic eNO levels in exhaled breath range from 20 to 80 ppbv.9,10 Therefore, NO detection and quantification at ppbv levels in human exhalation is an important medical diagnostic challenge.10
A number of different analytical methods, both optical and non-optical, have been developed to measure ultra low concentrations of various gases and, in particular, NO. Non-optical approaches include mass-spectrometry and gas chromatography. The main drawbacks of these techniques are size and cost of the apparatus, the need for sample conditioning and consumables, and the inability to make real-time online measurements. The most advanced optical techniques are based upon either chemiluminescence or laser absorption spectroscopy. Chemiluminescence is widely applied in measuring eNO levels from both children and adults. This is a FDA approved technique with a minimum detection sensitivity of < 1 ppbv.11 Two companies, Sievers and more recently Aerocrine, have developed analyzers based on a chemiluminescence approach for high precision real time NO monitoring. This technique requires calibration at the same humidity and temperature as breath and measures NO or NOx gases by reducing them to NO.
Another optical method for precise NO measurements is based on the Zeeman and Faraday effects, making use of the paramagnetic properties of NO. Background-free Faraday Laser Magnetic Resonance Spectroscopy (Faraday-LMRS) measures only nitric oxide in human breath.8,14,15 The method was evaluated in a clinical environment for on-line NO monitoring with a detection sensitivity of ~1 ppbv.8 However, the need for a magnetic field of more than 1000 Gauss and its inability to measure non-paramagnetic gases limits the applicability of Faraday-LMRS for clinical diagnostics. An ideal eNO sensor platform should be able to measure other gases in exhaled breath (for example, CO and CO2) for calibration and standardization purposes.16
In this work, we investigate tunable laser absorption spectroscopy (TLAS) as an effective technique for sensitive and selective NO monitoring. The mid-infrared spectral range is ideally suited for TLAS since most molecular gases possess strong fundamental rotational-vibrational lines in this region. The NO molecule has a strong absorption band around 5.2 mm (1900 cm-1) with a highest line intensity of ~ 6.04 × 10-20 cm-1/(molecule × cm-2) for the R(7.5) lines at 1903.1 cm-1. High resolution TLAS with a tunable single frequency mid-IR laser can resolve NO absorption features and selectively access NO spectral lines at low (= 100 Torr) pressure without interference from CO2 and H2O. This is particularly important in the development of biomedical gas sensors for breath analysis. Several types of CW spectroscopic laser sources operate in the mid-IR spectral region. These include the optical parametric oscillator, coherent sources based on different frequency generation, lead-salt diode lasers, and quantum cascade (QC) lasers. QC lasers were selected for this work because of their potential to meet the spectroscopic source requirements for compact, sensitive and selective gas detectors.17 Two continuous wave distributed feedback (DFB) QC lasers were utilized in our sensor architecture, as described in section 2. Each of these lasers has a spectral width of 3.0 MHz, and, together, they provide a total single mode tuning range of > 10 cm-1. In order to obtain detection sensitivities at ppbv or sub ppbv levels, long optical pathlengths (= 30 m), usually realized in multipass absorption cells, are used.16, 18-20 As can be deduced from the results presented by Nelson et al.,20 a sensitivity of 0.7 ppb/Hz1/2 can be achieved using a multipass cell with a 36 m path and a 0.3 liters volume (Aerodyne Research, Inc., Model AMAC 36).
Another approach for achieving long optical pathlength and minimizing the gas cell volume is the use of a cavity that consists of two ultra low-loss dielectric mirrors (~100 ppm) separated by a few cm. In this manner, effective optical pathlengths of hundreds of meters can be realized. Absorption in the cavity can be measured from the change in the cavity ringdown time or from the time-integrated cavity output. The first method, known as cavity ringdown spectroscopy (CRDS), has been successfully applied to measure NO concentration at ppbv levels.17,21
The second approach for measuring time-integrated cavity output is based upon the excitation of a dense spectrum of transverse cavity modes and time averaging of cavity output. In this scheme, the more modes are excited, the easier it is to suppress the related cavity throughput fluctuations. This method was first reported by two groups almost simultaneously and is referred to as integrated cavity output spectroscopy (ICOS) or cavity enhanced absorption spectroscopy (CEAS).22,23 ICOS is less technically demanding than CRDS, as it does not require fast time-resolved measurements and either high laser pulse energies or narrow cw laser linewidths. ICOS with an on-axis laser-cavity configuration has been used to measure NO with a detection limit of 16 ppbv.19 Off-axis ICOS (OA-ICOS) was proposed and implemented in the near IR because it provides increased spectral density of cavity modes and, thus, minimizes the noise in the resulting absorption spectra.24-26 In OA-ICOS the laser beam is directed off-axis with respect to the cavity axis. In this paper, we describe the use of the OA-ICOS technique with a CW mid-IR QC laser to measure NO concentrations in a compact absorption cell (76 cm3). A NO analyzer with a noise-equivalent (SNR = 1) sensitivity of 10 ppbv for OA-ICOS and 2 ppbv when using OA-ICOS with wavelength modulation was demonstrated and used to detect NO in nasal breath samples. The time required for data acquisition to obtain the reported sensitivity was 15 s in both cases.
Technical Plan and Equipment
The experimental arrangement is shown in Fig. 1. The laser source utilized in the
experiments is a continuous wave distributed feedback quantum cascade laser (CW DFB QC
laser) manufactured by Lucent Technologies Inc, operating at a single frequency at liquid
nitrogen (LN2) temperature.27 A laser chip containing several QC
lasers with output at ~ 5.2 mm (~1920 cm-1) is mounted on a cold finger inside a LN2
cryostat with one laser selected for usage. For better beam stability, a special design of
the LN2 cryostat (Cryo Industries Model # DET-1746-SLN) was used. The cryostat
has a rigid support for the LN2 reservoir, which prevents mechanical displacement of the
cold finger to which the QC laser is attached during the refilling process of the cryostat
and the entire LN2 holding period. A current driver (Wavelength Electronics,
MPL-5000) is used to operate the QC lasers. A current ramp with a maximum frequency of 5
kHz was applied to tune the laser frequency. We used two QC lasers in this work with
respective tuning ranges of 1911.51917.5 cm-1 and 1918.01923.0 cm-1.
The threshold current of each laser is ~ 300 mA and the maximum output power varies from
18 to 25 mW (depending on the wavelength and laser).

Figure 1. Off-Axis ICOS experimental arrangement; L1 and L2 are two collimating lenses, W1 and W2 are two 30 wedged uncoated ZnSe windows for frequency calibration, QCLquantum cascade laser inside the cryostat, and P is an off-axis parabolic mirror.
The QC laser radiation was collimated with two positive lenses (see Fig. 1). The first lens L1 is an aspheric ZnSe lens with an effective focal length (EFL) of 25.4 mm, a diameter of 27.94 mm, and broadband antireflection coatings (3-12 mm). The second lens L2 is a CaF2 bi-convex lens with a diameter of 12.7 mm and an EFL 13 mm. The collimator length is 130 mm, the distance between the QC laser output facet and L1 is 18 mm. Approximately 80 percent of the emitted power is collected assuming a divergence of 600 and a Gaussian beam envelope. The collimated beam diameter is about 3 mm and can be made divergent or convergent by changing the position of L2 relative to L1. This degree of freedom is important for optimal laser cavity mode coupling. Another advantage of the collimator is that by its use a simple routine is possible to perform optical alignment and realignment for different QC lasers on the same chip. A He-Ne laser is used for coarse alignment, since the 5.2 mm radiation is not visible. The final alignment of collimating lenses was performed with an IR camera (Thermovision).
The laser beam exiting the collimator is directed to the compact OA-ICOS cavity described in section 2 (see also Fig. 2). For spectroscopic measurements, laser radiation is scanned over a specific spectral range containing the gas absorption features of interest. Laser radiation is coupled into the cavity via the accidental coincidence of the laser frequency with a cavity mode.

Figure 2. Compact OA- ICOS Gas Cell
Radiation exiting the cell is focused onto a LN2 cooled photovoltaic HgCdTe detector (1 mm2 sensitive area) with a built-in 104 V/A transimpedance preamplifier (Kolmar Technologies, Model KMPV8-1-J1/DC). An off-axis parabolic mirror (3 inch in diameter and 3-inch focal length, 1 inch = 25.4 mm) is used to collect the transmitted cavity output, which was focused onto the detector area, thereby maximizing the light signal. A PC with a LabView 6.1 software based data acquisition and processing system is used for data accumulation, storage, and time averaging. Averaging of many laser frequency scans results in smoothing of the cavity resonance spikes.
The reflection coefficient or mirror losses at a specific wavelength can be estimated from the cavity decay time of the TEM00 mode, assuming that the reflectivity is uniform across the entire mirror surface.28 The intensity of a light pulse, trapped in an optical resonator, decreases exponentially with a time constant determined by the cavity finesse. In the case of a stable evacuated cavity formed by two identical mirrors and negligible Mie and Rayleigh scattering losses, the decay time %tempty for the TEM00 cavity mode or cavity ringdown time is defined only by the cavity length and the reflection coefficient (R) of the mirrors
| tempty = [L/c] [1 / (1 - R)] | (1) |
where L is the cavity length (5.3 cm) and c is the speed of light. For equation (1) we should consider the maximum experimental value of the decay time since the TEM00 mode has the lowest diffraction losses and, thus, the maximum ringdown time. We measured tempty = 0.7 ms, hence R 99.975%, or 250 ppm in terms of mirror losses. The QC laser linewidth has been estimated as described by Kosterev et al.21 From the cavity transmission resonance spikes, we obtain 2.9 MHz and 3.5 MHz for the linewidths of the two cw QC lasers used in this work.
The principal limitation to the detection sensitivity of ICOS is the cavity mode noise that results from incomplete averaging of the optical cavity transmission. This problem becomes more pronounced in the case of a short optical cavity. This can be explained by the larger FSR and, hence, wider gaps between the cavity transmission peaks, as well as higher mechanical stability of the short cavity. Our observations have shown that even for off-axis beam alignment, natural cavity instabilities like acoustic noise and temperature variations are not sufficient to average out the cavity mode structure. Therefore, in addition to laser scanning, the length of the cavity is dithered by PZT actuators at 200 Hz. The PZT mirror displacement amplitude corresponds to > 7 FSR of the modes resulting from off-axis alignment and, therefore, is sufficient to wash out the throughput spikes. The absorption signal is extracted from a measurement of the time-integrated light intensity that leaks out of the cavity.
A gas flow system enables us to control flow rate and perform NO concentration measurements at reduced pressures (in order to eliminate line broadening and minimize interference by other gases) and known flow conditions. A pressure controller can provide gas cell pressures ranging from 1 to 100 Torr. The gas flow system is also configured to permit off-line NO measurements of breath samples collected in special medical breath bags (Quintron, Model # QT00830-P).
Experimental Activity
According to the HITRAN database,29 the fundamental absorption band for NO is located in the region from
1780 cm-1 to 1950 cm-1. The optimum NO target wavelength is 1900.08 cm-1
for the R(6.5) component when taking into account maximum line intensity and minimum
interference from nearby absorption lines from other trace gas species.30 The
maximum intensity for R(6.5) lines is S = 6.0 × 10-20 cm-1/(molec × cm-2). However, the two QC lasers available for our
investigations cover the spectral ranges:
1)
1918.01923.0 cm-1 (QC laser #1), and
2) 1911.51917.5 cm-1 (QC Laser #2),
which do not overlap this line. Therefore a different choice of target wavelength was required.
The OA-ICOS technique was tested using two calibration mixtures of NO in N2 (Scott Specialty Gases, Inc) with known NO concentrations of 490 ppbv and 94.9 ppbv, respectively, at a pressure of 100 Torr. The NO concentration for the first mixture was independently measured using CRDS21 and, for the second calibration mixture, the NO concentration was determined by means of chemiluminescence. The time integrated transmitted intensity of the OA-ICOS cell filled with two NO-N2 calibration mixtures was measured for 5 × 104 averaged scans of the R(13.5) NO line at 1920.7 cm-1. The laser scan frequency was 3 kHz. The measured absorption spectra are in good agreement with the HITRAN simulation for an effective absorption pathlength of 80 m. An estimate of the detection sensitivity can be obtained from the deviation of the best-fit coefficients for a Voigt fit of the NO absorption line, which yields a noise equivalent sensitivity of 10 ppbv for 1 s deviation of the best-fit coefficient. Experimental results together with the fitting curve are given in Fig. 3.

Figure 3. Voigt Fit of the R(13.5) NO Absorption Line Data (same 95 ppbv NO in N2 mixture).
Feasibility experiments using OA-ICOS and wavelength modulation spectroscopy (WMS)31 were performed in order to determine biogenic NO concentrations from nasal exhaled air. The QC laser #2 current was scanned across the NO line by a triangular current ramp at a frequency of 8 Hz. A sinusoidal dither of f ~ 5 kHz (maximum for the laser current driver used in this work) was superimposed on the QC laser current ramp. The second harmonic (2f) signal of the OAICOS cavity output was sampled with a lock-in amplifier and averaged using a data acquisition card (DAQ Card 6062E, National Instruments) and LabView software. The amplitude of the 2f spectra is directly proportional to the NO concentration, which can be retrieved from calibration measurements. Fig. 4 depicts 2f spectra of the R(10.5) NO line at 1912.07 cm-1 for the 95 ppbv NO:N2 calibration mixture as well as for collected nasal NO. The nNO concentration was deduced to be 53 ppbv. It is apparent from these results (see Fig. 10 and Fig. 11) that wavelength modulation spectroscopy yields a better detection sensitivity than direct absorption spectroscopy. The increase of the SNR is > 5 and the noise equivalent detection sensitivity is ~ 2 ppbv. The SNR can be further improved by using a faster QCL current driver, which will enable us to achieve optimum values for the laser current ramp and sinusoidal dithering frequencies.

Figure 4. NO concentration measurement from nasal breath (1) using wavelength modulation applied to OAICOS. A NO Concentration of 53 ppbv was measured in nasal breath. (2) NO:N2 calibration mixture, 95 ppbv NO.
Conclusions
A CW QC laser-based NO sensor using a compact OA-ICOS cell has been developed, and its
performance characteristics have been investigated and compared with previous gas analyzer
designs based on CRDS and ICOS. The OA-ICOS-based sensor offers a similar noise equivalent
sensitivity of 10 ppbv as a CRDS based sensor platform, but is easier to align and more
robust in long term operation. Feasibility experiments of biogenic nNO concentration measurements from
the nasal cavity have been performed by adding wavelength modulation to improve the
detection sensitivity. A wavelength modulation spectroscopy noise equivalent sensitivity
of 2 ppbv has been achieved. The total data acquisition and averaging time was 15 s in the
both cases. Although this time exceeds a typical breath cycle duration, the time required
for a single data point measurement is much shorter (< 1 s). Therefore, the described NO measurement procedure can be
modified to measure NO concentration as a function of breath cycle phase. To achieve this
goal, primary data should be collected during several breath cycles and subsequently
sorted and averaged separately depending on the breath-cycle
phase.
Acknowledgments
We acknowledge the assistance of Dr. Anatoliy Kosterev in providing numerous useful
discussions and support. Financial support of the work performed by the Rice group was
provided by the Institute for Space Systems Operations (ISSO), the National Aeronautics
and Space Administration (NASA), the Texas Advanced Technology Program (TATP), the
National Institute of Health (NIH) via a sub contract from Physical Sciences, Inc., and
the Welch Foundation.
References
1M. W. Sigrist. Air Monitoring by Spectroscopic Techniques. New York: John Wiley & Sons, 1994.
2D. D. Nelson, J. L. Jimenez, G. J. McRae, M. S. Zahniser,
and C. E. Kolb. "Remote Sensing of NO and NO2 Emission from Heavy-Duty Diesel Trucks Using Tunable Diode
Lasers," in Application of Tunable Diode and Other Infrared Sources for
Atmospheric Studies and Industrial Processing Monitoring II. Ed. Alan Fried. Proc.,
SPIE 3758 (1999): 180-90.
3W. H. Weber, T. J.
Remillard, R. E. Chase, J. F. Richert, F. Capasso, C. Gmachl, A. L. Hutchinson, D. L.
Sivco, J. N. Baillargeon, and A. Y. Cho. "Using a Wavelength-Modulation Quantum
Cascade Laser to Measure NO Concentration in the Parts-per-Billion Range for
Vehicle Emissions Certification," Appl. Spectrosc. 56 (2002): 706-14.
4S. A. Kharitonov and P. J.
Barnes. "Clinical Aspects of Exhaled Nitric Oxide," Eur. Respir. J. 16
(2000): 781-92.
5S. A. Kharitonov and P. J.
Barnes. "Exhaled Markers of Pulmonary Disease," Am. J. Respir. Crit. Care Med.
163 (2001): 1693-722.
6B. Gaston, J. M. Drazen, J.
Loscalso, and J. S. Stamler. "The Biology of Nitrogen Oxide in the Airways," Am.
J. Respir. Crit. Care Med. 149 (1994): 538-51.
7K. Bhagat and P. Vallance.
"Nitric Oxide 9 Years Out," J. R. Soc. Med. 89 (1996): 667-73.
8P. Murtz, L. Menzel, W.
Bloch, A. Hess, O. Michel, and W. Urban. "LMR Spectroscopy: ANew Sensitive Method for
On-Line Recording of Nitric Oxide in Breath," J. Appl. Physiol. 86 (1999):
1075-80.
9D. H. Yates. "Role of
Exhaled Nitric Oxide in Asthma," Immunol. Cell Biol. 79 (2001): 178-90.
10L. Prieto.
"Measurements of Exhaled Nitric Oxide Concentration in Asthma. Technical Aspects and
Clinical Usefulness," Alergol. Immunol. Clin. 17 (2002): 72-87.
11N. Binding, W. Muller, P.
A. Czeschinski, and U. Witting. "NO Chemiluminescence in Exhaled Air:
Interference of Compounds from Endogenous or Exogenous Sources," Eur. Respir. J.
16 (2000): 499-503.
12H. W. Shin, C. M.
Rose-Gottron, R. S. Sufi, F. Perez, D. M. Cooper, A. F.Wilson, and S. C. George.
"Flow-Independent Nitric Oxide Exchange Parameters in Cystic Fibrosis," Am.
J. Respir. Crit. Care Med. 165.3 (2002): 349-57.
13M. Maniscalo, V. Di Mauro,
E. Farinaro, L. Carratu, and M. Sofia. "Transient Decrease of Exhaled Nitric Oxide
After Acute Exposure to Passive Smoke in Healthy Subjects," Arch. Environ. Health.
(2002) 57.5 437-40.
14G. Litfin, C. R. Pollock,
R. F. Curl, Jr., and F. K. Tittel. "Sensitivity Enhancement of Laser Absorption
Spectroscopy by Magnetic Rotation Effect," J. Chem. Phys. 12 (1980): 6602-05.
15H. Ganser, W. Urban, and
J. M. Brown. "The Sensitive Detection of NO by Faraday Modulation Spectroscopy
with a Quantum Cascade Laser," Molecular Physics 101 (2003): 545-50.
16C. Roller, K. Namjou, J.
D. Jeffers, M. Camp, A. Mock, P. J. McCann, and J. Grego. "Nitric Oxide Breath
Testing by Tunable-Diode Laser Absorption Spectroscopy: Application in Monitoring
Respiratory Inflammation," Appl. Opt. 41 (2002): 6018-29.
17A. A. Kosterev and F. K.
Tittel. "Chemical Sensors Based on Quantum Cascade Lasers," IEEE J. Quantum.
Electron. 38 (2002): 582-91.
18E. V. Stepanov, P. V.
Zyrianov, and V. A. Miliaev. "Single-Breath Detection with Tunable Diode Lasers for
Pulmonary Disease Diagnosis," in ALT98 Selected Papers on Novel Laser
Methods in Medicine and Biology. Eds. A. M. Prokhorov, V. I. Pustovoy, and G. P.
Kuzmin. Proc., SPIE 3829 (1999): 103-09.
19L. Menzel, A. A. Kosterev,
R. F. Curl, F. K. Tittel, C. Gmachl, F. Capasso, D. L. Sivco, N. J. Baillargeon, A. L.
Hutchinson, A. Y. Cho, and W. Urban. "Spectroscopic Detection of Biological NO with
a Quantum Cascade Laser," Appl. Phys. B 72 (2001): 859-63.
20D. D. Nelson, J. H.
Shorter, J. B. McManus, and M. S. Zahniser. "Sub-Part-Per-Billion Detection of Nitric
Oxide in Air Using a Thermoelectrically Cooled Mid-Infrared Quantum Cascade Laser
Spectrometer," Appl. Phys. B. 75 (2002): 343-50.
21A. A. Kosterev, A. L
Malinovsky, F. K. Tittel, C. Gmachl, F. Capasso, D. L. Sivco, J. N. Baillargeon, A. L.
Hutchinson, and A. Y. Cho. "Cavity Ringdown Spectroscopic Detection of Nitric Oxide
with Continuous-Wave Quantum-Cascade Laser," Appl. Opt. 40 (2001): 5522-29.
22A. OKeefe.
"Integrated Cavity Output Analysis of Ultra-Weak Absorption," Chem. Phys.
Lett. 293 (1998): 331-36.
23R. Englen, G. Berden, R.
Peters, and G. Meijer. "Cavity Enhanced Absorption and Cavity Enhanced Magnetic
Rotation Spectroscopy," Rev. Sci. Instrum. 69 (1998): 3763-69.
24J. B. Paul, L. Larson, and
J. G. Anderson. "Ultrasensitive Absorption Spectroscopy with a High-Finesse Optical
Cavity and Off-Axis Alignment," Appl. Opt. 40 (2001): 4904-10.
25D. S. Baer, J. B. Paul, M.
Gupta, and A. OKeefe. "Sensitive Absorption Measurements in Near-Infrared
Region using Off-Axis Integrated-Cavity-Output Spectroscopy," Appl. Phys. B 75
(2002): 261-65.
26V. L. Kasyutich, C. E.
Canosa-Mas, C. Pfrang, S. Vaughan, and R. P. Wayne. "Off-Axis Continuous-Wave
Cavity-Enhanced Absorption Spectroscopy of Narrow-Band and Broadband Absorbers Using Red
Diode Lasers," Appl. Phys. B 75 (2002): 755-61.
27F. Capasso, C. Gmachl, R.
Paiella, A. Tredicucci, A. L. Hutchinson, D. L. Sivco, J. N. Baillargeon, and A. Y. Cho.
"New Frontiers in Quantum Cascade Lasers and Applications," IEEE Select.
Topics Quantum Electron. 6 (2000): 931-47.
28J. M. Herbelin, J. A.
McKay, M. A. Kwok, R. H. Ueunten, D. S. Urevig, D. J. Spenser, and D. J. Benard.
"Sensitive Measurement of Photon Lifetime and True Reflectances in an Optical Cavity
by a Phase-Shift Method," Appl. Opt. 19 (1980): 144-47.
29L. S. Rothman, A. Barbe,
D. C. Benner, L. R. Brown, C. Camy-Peyret, M. R. Carleer, K. Chance, C. Clerbaux, V. Dana,
V. M. Devi, A. Fayt, J. M. Flaud, R. R. Gamache, A. Goldman, D. Jacquemart, K. W. Jucks,
W. J. Lafferty, J. Y. Mandin, S. T. Massie, V. Nemtchinov, D. A. Newnham, A. Perrin, C. P.
Rinsland, J. Schroeder, K. M. Smith, M. A. H. Smith, K. Tang, R. A. Toth, J. V. Auwera, P.
Varanasi, and K. Yoshino. "The HITRAN Molecular Spectroscopic Database: Edition of
2000 Including Updates through 2001," J. Quant. Spectrosc. Radiat. Transfer 82
(2003): 5-44.
30S. D. Wehe, D. M.
Sonnenfroh, M. G. Allen, C. Gmachl, and F. Capasso. "Quantum-Cascade Laser-Based
Sensor for CO and NO Measurements in Combustor Exhaust Flows," AIAA
2001-3317, 37 Joint Propulsion Conference and Exhibit, Salt Lake City, UT, July 8-11,
2001.
31D. S. Bomse, A. C.
Statlon, and J. A. Silver. "Frequency Modulation and Wavelength Modulation
Spectroscopies: Comparison of External Methods Using a Lead-Salt Diode Laser," Appl.
Opt. 31 (1992): 718-31.
Publications
Bakhirkin, Y. A., A. A. Kosterev, C. Roller, R. F. Curl, and F. K. Tittel.
"Mid-Infrared Quantum Cascade Laser-Based Off-Axis Integrated Cavity Output
Spectroscopy for Biogenic Nitric Oxide Detection," Appl. Opt. 43.11 (2004):
2257-66.
Kosterev, A. A., Y. A. Bakhirkin, F. K. Tittel, S. Blaser, Y. Bonetti, and L. Hvozdara.
"Photoacoustic Phase Shift as a Chemically Selective Spectroscopic Parameter," Appl
Phys B (2004). (Accepted for publication.)
Presentations
Bakhirkin, Y. A., A. A. Kosterev, R. F. Curl, F. K. Tittel, and M. Allen. "Off-Axis
Integrated Cavity Output Spectroscopy for Nitric Oxide Detection in Human Breath Using a
Quantum Cascade Laser," 87th OSA Annual Meeting, Tucson, AZ, Oct. 5-9, 2003.
Tittel, F. K., A. A. Kosterev, Y. Bakhirkin, C. B. Roller, D. Weidmann, and R.F. Curl.
"Quantum Cascade Laser Applications Chemical and Environmental Analysis,"
NEST-INFN & Scuola Normale Superiore Colloquium, Pisa, Italy, Nov. 7, 2003.
Bakhirkin, Y. A., A. A. Kosterev, C. Roller, R. F .Curl, and F. K. Tittel.
"Wavelength Modulation Off-Axis Integrated Cavity Output Spectroscopy for Biogenic NO
Detection in Human Breath," Conference on Lasers and Electro-Optics/International
Quantum Electronics Conference (CLEO/ IQEC 2004), San Francisco, CA, May 16-21, 2004.
McCurdy, M., Y. Bakhirkin, A. Kosterev, R. F. Curl, M. G. Allen, and F. K. Tittel.
"Off-Axis Integrated Cavity Output Spectroscopy for Nitric Oxide Detection in Human
Breath Using a Quantum Cascade Laser," AAAAI Meeting, San Francisco, CA, March 19-23,
2004.
McCurdy, M. R., Y. A. Bakhirkin, C. Roller, A. A. Kosterev, R. F. Curl, and F. K. Tittel.
"Off-Axis Integrated Cavity Output Spectroscopy for Nitric Oxide Detection in Human
Breath Using a Quantum Cascade Laser," 21st Annual Houston Conference on Biomedical
Engineering Research (HSEMB), Houston, TX, Feb. 12-13, 2004.
Tittel, F. K., Y. Bakhirkin, R. F. Curl, A. A. Kosterev, S. So, D. Weidmann, and
G.Wysocki. "Applications of Quantum Cascade Lasers in Trace Gas Analysis and Chemical
Sensing," 6th International Conference on Mid-Infrared Optoelectronics Materials and
Devices, St. Petersburg, Russia, June 28-July 1, 2004.
Tittel, F. K., A. A. Kosterev, Y. A. Bakhirkin, C. Roller, D. Weidmann, and R. F. Curl.
"Mid-Infrared Quantum Cascade Laser Sensors for Gas Analysis and Medical
Diagnostics," Photonics West 2004, San Jose, CA, Jan. 25-29, 2004.
Tittel, F. K., Y. Bakhirkin, R. F. Curl, A. A. Kosterev, M. McCurdy, C. Roller, S. So, D.
Weidmann, and G. Wysocki. "Current Applications of Quantum Cascade Laser in Trace Gas
Analysis and Chemical Sensing," International Workshop on Quantum Cascade Lasers,
Sevilla, Spain, Jan. 4-8, 2004.
Funding and proposals
"Photonic Technologies for Early Detection of Human Disease." NASA-National
Cancer Institute, Rice University PI, Apr. 1, 2002-Mar. 31, 2005, $377,000.
"Quantum Cascade Laser Based Sensors for Chemical and Environmental Analysis."
Texas Advanced Technology Program, Rice University PI., Jan. 1, 2002-Mar. 31, 2005,
$115,000.
PDF (423KB)
Table of Contents
Institute for Space Systems Operations - Y2003
Annual Report
Copyright © 2004
![]()
|